Transparent and translucent samples are not simply opaque samples with less color. Light can pass through them, scatter inside them and reflect from their surfaces. A repeatable method must therefore control the optical path, specimen position and the exact transmission or backed-reflectance result being reported.
Start with the measurement question
| Question | Likely method direction |
|---|---|
| What color is seen through the material? | Transmission color with a defined path length. |
| How much light passes through? | Luminous or spectral transmittance under a named method. |
| How cloudy is it? | A defined haze method, not L*a*b* alone. |
| How does it look over its final substrate? | Controlled backed reflectance may be appropriate. |
Control the sample and optical path
Thickness is part of the method. Two plaques made from the same resin can produce different transmission color when their path lengths differ. For liquids, use matched cells or cuvettes, keep the path length constant, remove bubbles and fill to a repeatable level. For a non-scattering solution, Beer–Lambert behaviour means absorbance scales with path length, while the resulting L*a*b* values do not scale linearly; a 20 mm cell result cannot be converted to a 10 mm result by simply halving the color coordinates. For films and sheets, control orientation, tension, surface side and the measured location.
Cleanliness is equally important. Fingerprints, dust, scratches and mould-release residue can change scattering and transmission. Handle specimens by the edges and define whether protective liners are removed. Temperature, settling and concentration can matter for liquid or dispersed samples.
Choose the optical result deliberately
Total transmission includes both directly transmitted and diffusely transmitted light when the instrument and method are configured to collect it. Regular transmission concentrates on the unscattered beam. Their difference helps explain why two samples can have similar color or total transmittance but different visual clarity. Instrument capability must match the specified method; a reflectance-only handheld cannot become a transmission instrument through software.
Blank and reference procedures matter
For liquids, the cell and solvent or carrier can contribute color and reflection. Define the reference used to zero or baseline the instrument, and keep it consistent between sites.
A routine measurement workflow
01
Condition and inspect
Check cleanliness, thickness, bubbles, damage and temperature before measurement.
02
Reference and position
Run the required baseline and place the specimen in the defined orientation and fixture.
03
Measure and document
Average representative readings and record mode, path length, settings and sample condition.
Pair this method with reflectance vs transmittance, sample preparation and a documented sampling plan.
Last reviewed 21/09/2026 · Level: Intermediate
Topics
Applications
Related knowledge
CIELAB
CIE L*a*b* Explained: Understanding L*, a* and b*
CIE L*a*b* describes color with three coordinates: L* lightness, a* red–green and b* yellow–blue. Learn how to interpret Lab values for practical quality control.
4–5 minute guide · Read the guide →Measurement Techniques
How to Measure Fabric Color
Measure fabric color with controlled conditioning, layers, backing, orientation, aperture and sampling so standards, lots and sites can be compared reliably.
Technical guide · Read the guide →Measurement Techniques
Measuring Yarn and Fibre Color
Measure yarn and loose fibre color with controlled packing, tension, mass, holder, backing and repacking so gaps and orientation do not dominate the result.
Technical guide · Read the guide →Measurement Geometry
Measuring Textured Fabrics
Measure textured fabrics by controlling pile state, pressure, aperture, geometry, orientation and spatial sampling across knits, terry, fleece and raised surfaces.
Technical guide · Read the guide →Measurement Geometry
Measuring Directional Fabrics
Measure directional fabrics by fixing warp, weft, nap or brushing orientation and deciding whether one viewing direction or a validated rotational average represents use.
Technical guide · Read the guide →UV Measurement
Measuring Fluorescent Textiles
Measure fluorescent and optically brightened textiles with a defined physical UV condition, textile reference standard, matched sample presentation and verified viewing light.
Technical guide · Read the guide →