Paint & Coatings

Measuring Glossy Coatings

Separate coating color from surface shine so a good Delta E does not hide a visible gloss, haze or levelling defect.

Technical guidePaint & Coatings

A glossy coating can pass a color tolerance and still look wrong. Specular gloss, reflection haze, distinctness of image and orange peel change the distribution and clarity of reflected light. A robust QC method measures coating color under a defined condition and controls the relevant appearance attributes separately.

Why glossy films are difficult

Small changes in levelling, cure, film thickness, polishing or contamination can alter the surface reflection without materially changing pigment color. Fingerprints and wiping marks are especially visible on dark high-gloss finishes. Curvature or poor contact at the measurement port can introduce additional variation.

Choose the color condition

SCI includes specular reflection and reduces the influence of gloss differences. It is often the clearer choice for pigment, tinting and batch color. SCE excludes specular reflection and responds more strongly to the as-seen finish. Choose one for pass/fail from the product requirement. Where an instrument reports both, use the secondary mode as diagnostic information rather than silently switching between modes.

Measure gloss separately

Use a glossmeter and the geometry required by the specification. ASTM D523 covers 20°, 60° and 85° geometries for nonmetallic specimens. A 60° result is commonly used as the general measurement, with 20° or 85° providing better discrimination at the high- or low-gloss ends when the applicable method calls for them. Keep geometry fixed: values from different gloss angles are not interchangeable.

If the complaint concerns milky reflections, poor image clarity or orange peel, ordinary specular gloss may not explain it. Add reflection-haze, DOI or texture assessment when those properties drive acceptance.

Control the specimen

  • Condition standard and sample to the same temperature and cure age.
  • Handle by edges and use an agreed cleaning procedure that does not polish or swell the film.
  • Measure flat areas large enough for the color aperture and glossmeter footprint.
  • Keep application direction and instrument orientation fixed.
  • Take multiple readings across representative production locations.

Diagnose color–appearance disagreements

Result pattern First checks
SCI stable; SCE and gloss move Levelling, cure, film build, polishing, contamination and surface texture.
SCI and SCE move together Pigment, tint strength, dispersion, substrate show-through and formulation.
Gloss stable; visual image looks hazy Reflection haze, DOI, microtexture and orange peel.

Link the result to SCI vs SCE for coatings, paint tolerances and the complete coatings guide.

Last reviewed 22/09/2026 · Level: Intermediate

Topics

Applications

Related knowledge

Paint & coatings color measurement support

Need a method matched to your coating?

Tell us the coating, sample form, finish and tolerance. We’ll help identify a practical instrument and measurement workflow.